10-4
Troubleshooting Chart
10.2
SYMPTOM:
MEASUREMENT ERROR: IB Level Error
POSSIBLE CAUSE:
Pinched, leaking or disconnected tube.
ACTION:
Fix or replace tubing.
SECTION:
POSSIBLE CAUSE:
Sipper misaligned.
ACTION:
Check Sipper alignment.
SECTION:
POSSIBLE CAUSE:
Sipper pump not operating properly.
ACTION:
Replace sipper pump seals.
SECTION:
8.4.4 Sipper pump Seal Replacement
POSSIBLE CAUSE:
Stir bar not in sample module.
ACTION:
Install stir bar.
SECTION:
POSSIBLE CAUSE:
Stir speed too fast or too slow.
ACTION:
Adjust stir speed.
SECTION:
POSSIBLE CAUSE:
Newly installed enzyme Membrane.
ACTION:
Enter probe diagnostics and check probe currents.
SECTION:
POSSIBLE CAUSE:
Newly installed probe.
ACTION:
Enter probe diagnostics and check probe currents.
SECTION:
POSSIBLE CAUSE:
Power disruption.
ACTION:
Enter probe diagnostics and check probe currents.
SECTION:
POSSIBLE CAUSE:
Failing enzyme membrane.
ACTION:
Perform daily operational checks and replace membrane
if necessary.
SECTION:
5.1 Perform Daily Operational Checks, 4.7 Install Enzyme Membranes
POSSIBLE CAUSE:
Enzyme Probe surface fouled.
ACTION:
Clean probe surface.
SECTION:
POSSIBLE CAUSE:
Auxiliary electrode fouled.
ACTION:
Clean Auxiliary electrode surface (temperature probe) with isopropyl alcohol.
POSSIBLE CAUSE:
Sample may contain an interfering substance.
ACTION:
Attempt to confirm interference.
SYMPTOM:
MEASUREMENT ERROR: PL Limit Error
POSSIBLE CAUSE:
Sipper misaligned.
ACTION:
Check Sipper alignment.
SECTION:
POSSIBLE CAUSE:
Stir bar not in sample module.
ACTION:
Disassemble sample module and reinstall stir bar.
SECTION:
Содержание YSI 2500
Страница 1: ...USER MANUAL 525021 YSI 2500 Biochemistry Analyzer OPERATIONS AND MAINTENANCE MANUAL...
Страница 34: ...5 11 8 The Stat sample results are displayed on the Run Stat tab Stat sample results...
Страница 88: ...13 1 13 Notices Declaration of Conformity 13 1...
Страница 96: ...17 1 17 Appendix D Line Power Cord and Plug Wiring United Kingdom United States Europe...