3-3
Display
Graphical color LCD covered by a touch screen
USB ports
The USB ports allow a flash drive to be connected to the 2500
Analyzer to download sample results or upgrade the instrument’s
software. A USB port is located on the right side of the display. An
additional USB port is located on the rear of the instrument.
Sipper
Can be raised, lowered, rotated, and moved horizontally to its
various positions.
The positions are: Calibrator Well, Sample Module, and Stations 1
and 2.
The Sipper senses fluid level to control immersion depth and detect
errors.
Station 1
Plate and rack holder accepts most standard plates/racks for batch
sampling of up to 96 samples.
Station 2
Test tube holder for manual sampling.
Buffer Pump
Draws buffer from the buffer bottle, pumps it through the Sipper
Pump and the Sipper, and flushes the Sample Module.
Calibrator Pump
Draws the standard solution from the Calibrator Bottle and fills the
Calibrator Well.
Figure 3-4
Sipper pump
It retracts its piston to draw standard from the Calibrator Well or
sample from the sample stations.
It extends its piston to dispense standard or sample into the sample
module.
Sample Module
Sipper Pump