![Xylem WTW Multi 9430 Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/xylem/wtw-multi-9430/wtw-multi-9430_operating-manual_4023879094.webp)
Settings
Multi 9430
94
ba75966e04
04/2015
Setting menu of
LR 925/01
Menu item
Possible
setting
Explanation
Cell constant
0.090 ... ...
0.100
...
0.110 cm
-1
Display and setting options for the
cell constant
Temp. comp. (TC)
/
Method
nLF
lin
off
Procedure for temperature compen-
sation (see section 8.2 T
This setting is only available for the
measured parameters,
ϰ
and
ρ
.
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
3.000 %/K
Coefficient of the linear temperature
compensation.
This menu item is only available
when the linear temperature com-
pensation is set.
Temp. comp. (TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters,
ϰ
and
ρ
.
Multiplier for TDS
0.40 ...
1.00
Factor for TDS value
Reset
-
Resets all sensor settings to the
delivery condition (see section 9.7.1
R