Multi 9430
Dissolved oxygen (D.O.)
ba75966e04
04/2015
75
Proceed as follows to calibrate the instrument:
Before calibrating via a comparison measurement, the sensor
should be calibrated in the air calibration vessel.
1.
Connect the D.O. sensor to the meter.
2.
Immerse the D.O. sensor in the reference solution.
3.
Start the calibration with
<CAL>
.
4.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the end of the AutoRead measurement or accept the calibration
value with
<ENTER>
.
The factor that was set last is displayed.
6.
Using
<
> <
>
, set the correction factor to adjust the displayed con-
centration value to the nominal value (value of the comparison mea-
surement). Subsequently, accept the correction factor with
<ENTER>
.
The meter switches to the measured value display.
The status display
[Factor]
is active.
03.04.2013 08:00
03.04.2013 08:00