Settings
Multi 9430
90
ba75966e04
04/2015
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration
record.
Further options:
Scroll through the cali-
bration records with
<
><
>
.
Output the displayed cal-
ibration record to the
interface with
<F2>
/
[USB-Ausgabe]
.
Output all calibration
records to the interface
with
<F2_>
[USB-Aus-
gabe]
.
Quit the display with
<F1>
/
[Back]
or
<ENTER>
.
Switch directly to the
measured value display
with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to RS232/
USB
-
Outputs the calibration
records to the interface.
Man. temperature
-25 ...
+25
... +130 °C
Entry of the manually deter-
mined temperature. For
measurements without tem-
perature sensor only.
Alternative tempera-
ture
on
off
Takes the measured tem-
perature value from an IDS
sensor.
This menu item is only avail-
able if an IDS adapter and
an IDS sensor with temper-
ature sensor are connected.
Temperature from
channel
▭▭▬
▭▬▭
▬▭▭
Selection of the channel
from which the temperature
value should be taken over.
This menu item is only avail-
able if two IDS sensors are
additionally connected.
ISE setup
/
AutoRead criterion
low
medium
high
Selection of the AutoRead
criteria (see section 6.1.1
M
Menu item
Possible setting
Description