Settings
Multi 9430
86
ba75966e04
04/2015
9.1.2 Buffer sets for calibration
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.
Reset
-
Resets all sensor settings to the delivery condition (see
section 9.7.1 R
Menu item
Possible
setting
Explanation
No.
Buffer set *
pH values
at
1
ConCal
Any
Any
2
NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers
1.679
4.006
6.865
9.180
12.454
25 °C
3
TEC
WTW Technical buffers
2.000
4.010
7.000
10.011
25 °C
4
Merck 1*
4.000
7.000
9.000
20 °C
5
Merck 2 *
1.000
6.000
8.000
13.000
20 °C
6
Merck 3 *
4.660
6.880
9.220
20 °C
7
Merck 4 *
2.000
4.000
7.000
10.000
20 °C
8
Merck 5 *
4.010
7.000
10.000
25 °C
9
DIN 19267
1.090
4.650
6.790
9.230
25 °C