Ion concentration
Multi 9430
56
ba75966e04
04/2015
6.3
Selecting the measuring method
The following methods are supported:
Standard addition
Standard subtraction
Sample addition
Sample subtraction
Blank value addition
1.
Connect the ISE combination electrode to the meter.
The pH/ISE measuring window is displayed.
2.
If necessary, select the ISE display with
<M>
.
3.
If necessary, measure the temperature of the test sample with a ther-
mometer.
4.
Open the ISE menu with
<ENTER>
.
5.
Thoroughly rinse the electrode with distilled water.
6.
Adjust the temperature of the standard solutions.
7.
Select
Method
with
<
> <
>
and confirm with
<ENTER>
.
8.
Select a method with
<
> <
>
and confirm with
<ENTER>
.
9.
Select
Start method
with
<
> <
>
and confirm with
<ENTER>
.
Measurement with the selected method begins (see section 6.3.1
S
Calibration
Man. temperature
:
25 °C
ISE setup
Method
:
Standard addition
Start method
ISE
01.08.2013 08:00
Содержание WTW Multi 9430
Страница 1: ...OPERATING MANUAL ba75966e04 04 2015 Multi 9430 DIGITAL METER FOR IDS SENSORS...
Страница 113: ...Multi 9430 Maintenance cleaning disposal ba75966e04 04 2015 113...
Страница 131: ......