SonTek, a Xylem brand
130
RiverSurveyor S5/M9 System Manual (February 2013)
Figure 113. Site Information
G-7.4 Step 2: System Settings
Click
System Settings
to input Tagline Azimuth, Salinity, Track Reference, and Depth Refer-
ence. This will display the System Settings dialog (
114). Just like the Site Information di-
alog, press the
Tab
key to step through each of the displayed parameters.
Figure 114. System Settings
Tagline Azimuth: It is important to set the tagline azimuth if you are using the ENU orientation.
The tagline azimuth (
115) is the heading (direction) in degrees from magnetic north of
the tagline going towards the right bank. The normal velocity is used to compute discharge for
each station. The following illustration shows the proper orientation of the M9/S5 (connector
should always be oriented pointing downstream), the tagline azimuth relative to magnetic north
and the REW, and the downstream flow direction.
Содержание SonTek River Surveyor M9
Страница 10: ......
Страница 118: ......
Страница 150: ...SonTek a Xylem brand 140 RiverSurveyor S5 M9 System Manual February 2013 Figure 126 Island Edge...