60
6.2 Error messages
•
Error while measuring
Error code
Display
message
Errors
Description
Remedy
0x02000201
Timeout at
upper LB
Timeout at upper
light barrier
(A measurement was
started; no meniscus
passed through the
upper light barrier
within the timeout
period of 120
seconds)
1. Viscometer is not
positioned properly.
2. The sample liquid is
not transparent enough.
1. Check the position of
the viscometer.
2. Sample cannot be
measured!
Contact the
Service Center (see
back of
0x02000203
Detection
Error
Wrong signal on the
lower light barrier
(The measurement
was started and no
meniscus passing
was detected on the
lower light barrier
before one was
detected on the
upper light barrier)
1. Air bubbles in the
sample liquid.
2. Sample was not
above the upper light
barrier at the time of
triggering.
1. Air bubbles in the
sample liquid.
2. Pump the sample
above the upper light
barrier prior to starting
the measurement.
•
Error while saving the measuring results
Error code
Display message
Errors
Description
Remedy
0x02000301
Low internal
memory
Internal data
memory almost
full
Note: The internal
memory has been
populated almost
entirely with measured
values.
Save, print or delete
data
0x02000302
Internal memory
full
Internal data
memory full
All memory slots
populated with
measured values. New
measured values cannot
be saved
Save, print or delete
data
0x02000311
…12
CSV R/W Err
on USB OTG
Error during
writing into CSV
file to USB OTG
or USB host
The creation or writing of
the CSV file at the
respective USB interface
has failed
Reconnect the USB
drive or restart the unit.
Format the USB drive if
necessary (FAT16/32)