Conductivity
Lab 955
18
ba77144e02
12/2017
The stability control function (
AutoRead
) continually checks the stability of the
measured values in the monitored time interval. The stability has a consider-
able impact on the reproducibility of measured values. The
[AR]
display indi-
cator flashes until a stable value is measured.
Stability criteria (AutoRead )
5.1.2
Measuring the temperature
The temperature measurement is absolutely essential for a reproducible
conductivity measurement.
You have the following options to measure the temperature:
• Automatic measurement of the temperature with the temperature sensor
(NTC30 or Pt1000) integrated in the sensor.
• Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
Which type of temperature measurement is active is indicated by the tempera-
ture display and the
[TP]
status indicator:
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
1.
Measure the current temperature of the test sample.
2.
Set the temperature value with
<
><
>.
Measured
parameter
Time inter-
val
Stability in the time interval
Conductivity
ϰ
10 seconds
∆
: better than 1.0 % of measured value
Temperature
15 seconds
∆
: Better than 0.5 ° C
Tempera-
ture sensor
Resolution of
the temp. dis-
play
Status indica-
tor
Temp. measure-
ment
yes
0.1 °C
[TP]
Automatic with tem-
perature sensor
-
1 °C
-
Manual