Lab 955
Conductivity
ba77144e02
12/2017
17
5
Conductivity
5.1
Measuring
5.1.1
Measuring the conductivity
1.
Connect a conductivity measuring cell to the measuring instrument.
The conductivity measuring screen is displayed.
2.
Adjust the temperature of the solutions and measure the current temper-
ature if the measurement is made without a temperature sensor.
3.
Immerse the conductivity measuring cell in the test sample.
4.
When measuring without temperature sensor: Enter the temperature of
the test sample with
<
><
>.
5.
Wait for a stable measured value.
The
[AR]
display indicator no longer flashes.
6.
If necessary, select the measured parameter with
<MODE>
.
•
Conductivity [µS/cm] / [mS/cm]
•
Salinity Sal
•
Total dissolved solids TDS [mg/l]
If necessary, set the TDS factor with
<
><
> (see section 6.1.4
•
Resistivity [M
Ω
·cm]
Stability control (AutoRead )
During the measuring procedure, the stability control function is automatically
activated.
To ensure the high measurement accuracy of the measuring sys-
tem, measure with a current calibration or the correctly set cell
constant (see section 5.2 C
0
5
2
0
3
5
C
°
S/cm
m
nLF
TP
Tref25
AR
ARng