Lab 955
Reset
ba77144e02
12/2017
29
Measurement and system settings that can be reset
The following settings are reset to the delivery condition
(default):
• Measurement settings
• System settings
• Calibration data
The measuring system is possibly not calibrated after a reset.
Before measuring, make sure the meter uses the cell con-
stant suitable for the measuring cell.
Measurement settings
Default
Reference temperature (
Tref
)
t25
Unit of the measured temperature
value (
Uni
)
°C
Calibration interval (
Int.C
)
180 d
Temperature compensation
nLF
Temperature coefficient of the linear
temperature compensation
2.000 %/K
TDS factor
1.00
Manually set cell constant
0.475 cm
-1
System settings
Default
Switch-off interval (
.Off
)
1 h