What to do if...
Lab 955
32
ba77144e02
12/2017
9
What to do if...
9.1
Conductivity
9.1.1
Error message E3
9.1.2
Error message OFL, UFL
The measured value is outside the measuring range.
9.2
General information
9.2.1
[LoBat]
Display
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
• Measuring cell contaminated
• Clean the measuring cell and
replace it if necessary
• Calibration solution not suitable
• Check the calibration solutions
Cause
Remedy
• Measured value outside the
measuring range
• Use suitable conductivity
measuring cell
Cause
Remedy
Batteries almost empty
Replace the batteries (see section
3.3.1 I