Lab 955
Settings
ba77144e02
12/2017
25
6.1.2
Cleaning interval
The cleaning interval is set with the measurement settings (see section 6.1
M
The calibration evaluation (sensor symbol) flashes when a valid calibration is
available for a measuring cell and the set cleaning interval has expired.
The flashing calibration evaluation reminds you to clean and calibrate the
measuring cell regularly (see section 5.2.3 D
It is still possible to measure.
6.1.3
Selecting the temperature compensation
1.
Open the calibration menu with
<CAL>
.
2.
If necessary, select the
Use tC
setting with
<CAL>
.
3.
Press
<MENU/OK>
or
<CAL
_
>
to confirm the selection of
Use tC
.
The set temperature compensation is displayed.
4.
If necessary, change the setting of the temperature compensation with
<CAL>
.
•
nLF
: Nonlinear temperature compensation
•
Lin
: Linear temperature compensation
If necessary, set the linear temperature coefficient with
<
><
>.
•
----
: No temperature compensation (temperature compensation
switched off)
5.
If necessary, quit the setting of the temperature compensation with
<MENU/OK>
.
The set temperature compensation is used.
Temperature compensation
The calculation of the temperature compensation is based on the preset refer-
ence temperature, 20 °C [Tref20] or 25 °C [Tref25].
F
L
n
0
3
5
S/cm
m
nLF
Tref25
ARng