51
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
EN 5.10 SuSix Manual 2007
Data Logger
SuSix®
provides a data logger with a capacity of approx. 20,000 entry
points. See Appendix D on page
for examples and descriptions of log
files.
The data logger operates after the FIFO principle (First In, First Out). If the
data logger is full and new data are coming in, the oldest data are dis-
carded.
1. Press the up/down keys to highlight the required menu line
(here: Data Logger) and then press OK to select it.
Specify Main Screen
Data Logger
Password
Set Sensor Name
Converter Setup
Display Setup
Back Select
OK
Login
Activate/deactivate
Change password
Back Select
OK
Set log interval
0010 sec
Back Select >
Data Logger
Set Sensor Name
SuSix1
Back Select
>
Set Sensor Name
Select line
Linie 1 - Sensor name
Linie 2 - Meas.
Linie 3 -
Linie 4 -
Linie 5
Back Select
OK
Specify Main Screen
Converter Setup
Language
Set Clock
Modbus
Factory settings
Display SW version
Back Select
OK
Display Setup
Sensor Calibration
Averaging
Units
mA output
Cleaning
Solid curve
High Alarm
Low Alarm
Sensor error
Status
Factory settings
Service Menu
Back Select
OK
Select line content
Not in use
Meas.
Homogeneity
Solid curve
Clock
Back Select
OK
Password
Type password
0000
Back Select
>
Change password
0000
Back Select
>
Activate
Deactivate
Back Select
OK
Login
Activate/
Deactivate
Change password
SuSix1 [X]
Meas.
FTU
Select Setup
Main Menu
Converter Top Display
SuSix1 [X]
Meas.
FTU
Select Setup
Converter Top Display
Specify Main Screen
Data Logger
Password
Set Sensor Name
Converter Setup
Display Setup
Back Select
OK
Main Menu
SuSix1 [X]
Prim.
mg/l
Second 0.1%
Select Setup
1 - Main menus
Page1
2. Press the up/down keys to change the highlighted digit, and then press
the right-key (>) to proceed to the next digit.
Specify Main Screen
Data Logger
Password
Set Sensor Name
Converter Setup
Display Setup
Back Select
OK
Login
Activate/deactivate
Change password
Back Select
OK
Set log interval
0010 sec
Back Select >
Data Logger
Set Sensor Name
SuSix1
Back Select
>
Set Sensor Name
Select line
Linie 1 - Sensor name
Linie 2 - Meas.
Linie 3 -
Linie 4 -
Linie 5
Back Select
OK
Specify Main Screen
Converter Setup
Language
Set Clock
Modbus
Factory settings
Display SW version
Back Select
OK
Display Setup
Sensor Calibration
Averaging
Units
mA output
Cleaning
Solid curve
High Alarm
Low Alarm
Sensor error
Status
Factory settings
Service Menu
Back Select
OK
Select line content
Not in use
Meas.
Homogeneity
Solid curve
Clock
Back Select
OK
Password
Type password
0000
Back Select
>
Change password
0000
Back Select
>
Activate
Deactivate
Back Select
OK
Login
Activate/
Deactivate
Change password
SuSix1 [X]
Meas.
FTU
Select Setup
Main Menu
Converter Top Display
SuSix1 [X]
Meas.
FTU
Select Setup
Converter Top Display
Specify Main Screen
Data Logger
Password
Set Sensor Name
Converter Setup
Display Setup
Back Select
OK
Main Menu
SuSix1 [X]
Prim.
mg/l
Second 0.1%
Select Setup
1 - Main menus
Page1
3. Continue until all digits have been set and press OK.
The log interval can be set in intervals from 10 seconds up to 9999
seconds.