33
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
EN 5.10 SuSix Manual 2007
7. See on the following picture where two jumpers must be inserted for a
115 V AC configuration:
115 V AC dual jumper positions
8. Insert and solder 2 jumpers (wires) in the positions indicated by the
black arrows.
9. Turn around the PCB and replace the 63 mAT fuse on the right side
with a 125 mAT fuse ( or vice versa going from 115 VAC to 230 VAC).
10. Re-insert the PCB and re-assemble the converter (see also steps 4, 3,
2 and 1).