Chapter 3 Controller Descriptions and Operation
3
[EasyAccess]: EasyAccess is a value-added service provided by the manufacture,
by use of internet and PC or mobile clients based on Penetration communication. For access
of the function, contact Xtemp or your local engineer for information, activation and
payment.
[Fan Speed]: Used to set the speed of the circulating fan or circulating pump. Optional for
specific equipment only.
[Dehmd Comp]: Used to set the parameters of a compressor for dehumidification. Optional for
specific equipment only.
[System Para]: Settings for system parameters including password for unlock of the HOME screen;
password, IP address and gateway for remote control VNC, equipment code, taskbar, key
tone, mouse, brightness and backlight delay etc, where VNC applies only to equipment
incorporating Ethernet (Fig 36).
Where backlight is required all the time, input 0 for continuous backlight.
Fig 36
[Heater Para]: Used to set additional heating once T/H MODE set to stepper PID, refer to 3.15.
[T/H MODE]: Used to set PID mode for better temperature and humidity performance, refer to 3.15
for details.
[RADIATION SET]: Used to set the dimming rate of the lights from one level to another, and safe
radiation level for the lamp as operator light or protection against optical hazard from door
operation. Refer to 3.16 for details.
3.15 T/H MODE
At screen of USER, tap [T/H MODE] to enter screen of T/H MODE for stepper PID or heater PID (Fig
37). There are 2 modes for selection, MODE 0 or MODE 1,