XTEMP ENVIRONMENTS (SHANGHAI) CO., LTD.
http://www.xtemp-usa.com/
9
/
55
DSEO0711-2-V211119/UXP8401-PG.CC.GE.AR-CD7003/5~+45
℃
Chapter 1 Safety Warning and Instructions
1
Connect the power from behind the equipment. Plug in and remove the cable by directly
holding the plug. No dragging of cable in any part. Protect the cable from being damaged by
contacting with the hot surface of the equipment or mouse etc!
Keep clear of inlets and outlets of back plenum for the circulating hot air. Excessive hoarding of
specimen or blockage of the air circulating may cause frost accumulation on surface of the
evaporator, localized temperature distribution and degraded performance!
Install a flexible tubing to the extraction system, if the operating of the equipment could lead to
liberation of hazardous air or gas mixture, for example the application of chemicals to treat
specimen under growth!
Do not disassemble or exchange assembly parts and electric circuits by unauthorized
personnel! Hazard of possible electric shock! In case of repairing, contact authorized local
distributor or maintenance engineer.
Always remove the mains plug from the socket before cleaning, moving, maintenance or
repairing!
If you come in touch with the PCB's of the equipment please take care that you follow the
electro-static discharge (ESD) guidelines and use ESD protection kit to prevent any damage on
the electronic components.
Perform regular check or maintenance for the following at designated internals:
Clean the cabinet interior weekly or where necessary!
Decontaminate cabinet interior and wire shovels on case-by-case bases if necessary!
Check the air tightness of the door seal at six months internals!
Check and clean the condenser at six months internals or where necessary!
Check cabinet Klixon at six months intervals or more frequently when necessary!
Check low level cut-out switch LLCO for the humidifier at three months intervals or as
necessary!