XS Scuba Airwave Octopus Скачать руководство пользователя страница 18

AirWave Octo

18

AirWave Octo Service and Repair Manual

slightly counter-clockwise while lightly depressing 

the purge cover to prevent wear on the seat.  Do not 

adjust any further than is needed to establish a slight 

leak.

6.  Hold the drive of the Inline Adjustment Tool engaged with 

the orifice.  While lightly depressing the purge cover, turn 

the orifice slightly clockwise a very small fraction of a turn.  

Pause after each adjustment to listen, and be careful to 

avoid over-adjusting beyond the point that the leak has 

stopped.  When the leak has stopped, purge the second 

stage again to ensure that it does not return.  Observe 

the IP Test Gauge while purging to verify that it indicates 

a stable intermediate pressure after each cycle, with no 

creep or fluctuation.

CAUTION:  Over-adjustment of the orifice can cause 
excessive spring load in the second stage valve, and 
may impair the regulator’s performance.

8.  When the second stage is properly adjusted with no leaks 

or lever slack, depressurize and purge the system to dis-

connect the LP hose.

9.  Reconnect the LP hose to the inlet coupling, and apply a 

torque wrench with 17mm crow-foot to tighten the  fitting 

to a torque measurement of 55 inch-lbs (±5).

Subjective Breathing Test

1.  Connect the regulator to a cylinder containing 2,500 – 

3,000 psi, and open the valve to pressurize the system.

2.  Turn the venturi switch counter-clockwise to the open or 

"dive" position.

3.  Fully depress the second stage purge to ensure that an 

adequate volume of air flows through the mouthpiece, 

sufficient to clear the second stage.  Then, breathe sev

-

eral times from the second stage.

A properly serviced and adjusted regulator should de-

liver air upon deep inhalation without excessive inhala-

tion effort, freeflow, or vibration.  When exhaling, there 

should be no resistance or sticking of the exhalation 

valve.  If any of these problems occur, refer to 

Table 

2 - Troubleshooting. 

Содержание Airwave Octopus

Страница 1: ...Service Repair Manual rev 06 08 SERVICE AND REPAIR MANUAL AIRWAVE OCTO PN RG320...

Страница 2: ...External Inspection 6 Immersion Leak Test 6 Intermediate Pressure Test 7 Section 3 Disassembly Procedures General Guidelines 8 Section 4 Reassembly Procedures General Guidelines 12 Section 5 Final Tes...

Страница 3: ...product it is extremely critical that it receives service according to the procedures outlined in this manual on a regularly scheduled basis at least once a year with normal or infrequent use NOTE A...

Страница 4: ...sult from improper use of this tool Be sure to follow the instructions provided in this manual NOTE XS Scuba especially endorses and recommends the Jaws Gripmaster Bench Vise available from Peter Buil...

Страница 5: ...ol PN 20 500 200 Tech Driver IPG PN 20 165 111 O ring Tool Set PN 10 125 400 1 4 Slotted Screwdriver PN 11 465 500 1 4 Nut Driver PN 11 168 500 Small 02 Blade Screwdriver PN 11 133 500 Cristo Lube PN...

Страница 6: ...r damage that exposes the brass beneath the finish CAUTION Do not clean any parts that show damage to their finish inside an ultrasonic cleaner 4 Closely examine all parts of the first stage for any o...

Страница 7: ...xceeds 400 psi may result in damage or rupture of the test gauge or LP hose and could cause serious personal injury 2 Slowly open the supply valve to pressurize the first stage Closely monitor the IP...

Страница 8: ...ngs After passing close inspection static O rings may sometimes be reused although this is not necessarily recommended Dynamic O rings must be automatically discarded and replaced with every service r...

Страница 9: ...a n open end wrench turn the inlet fitting counter clockwise to loosen and unscrew it completely by hand to remove Closely examine the plastic threads and overall condition of the inlet fitting and re...

Страница 10: ...tor counter clockwise from the air barrel 15 and carefully remove the O ring 21 with a plastic tool 13 Turn the air barrel over to allow the balance shuttle cap 20 and LP spring 19 to drop out see Fig...

Страница 11: ...ace to keep it isolated from other metal parts 21 Gently pull the exhaust deflector 26 off from the second stage housing to expose the exhaust valve 25 for inspection 22 Closely inspect the exhaust de...

Страница 12: ...G 11 PN MS150 to be used exclusively for the lubrication and dressing of O rings and other internal parts Christo Lube provides superior lubrication to that of silicone grease especially in high press...

Страница 13: ...e tab engages in the hole next to it Then gently press the opposite leg into place so the the other tab engages in the opposite hole When done cor rectly the lever pad should rest flush against the fl...

Страница 14: ...te leg engages with the opposite side The lever should now be standing up when released pointing toward the venturi deflector See Fig 14 11 While holding the second stage housing right side up with th...

Страница 15: ...t clockwise while closely watching the lever When the lever starts to drop stop do not turn the orifice any further This will indicate that the orifice has made contact against the LP seat See Fig 20...

Страница 16: ...ate pressure exceeds 155 170 psi ensure that a fully assembled and properly adjusted second stage is connected to the first stage before pressurizing Failure to relieve intermediate pressure that exce...

Страница 17: ...justing Tool and mate the opposite fitting of the tool onto the inlet fitting of the second stage Pull back the knurled knob to retract the slotted drive and tighten the tool clockwise onto the inlet...

Страница 18: ...stage valve and may impair the regulator s performance 8 When the second stage is properly adjusted with no leaks or lever slack depressurize and purge the system to dis connect the LP hose 9 Reconne...

Страница 19: ...e to slower bubble formation that occurs in smaller leaks Disassemble the regulator at the source of the leak to check sealing surfaces assembly sequence and component positioning in order to correct...

Страница 20: ...ermediate pressure 1 Refer to first stage troubleshooting 2 Orifice incorrectly adjusted 2 Reset to preliminary setting and readjust 3 Lever is damaged 3 Replace with new 1 Lever is slack or orifice i...

Страница 21: ...rifice 10 RP320 10 1 O ring 11 RP320 11 1 Mouthpiece 12 RP320 12 1 Tie Strap 13 RP320 13 1 Housing 14 RP320 14 1 Lever Pad 15 RP320 15 1 Air Barrel 16 RP320 16 1 LP Seat 17 RP320 17 1 Poppet 18 RP320...

Отзывы: