PIXIE-4 User’s Manual
V2.69
©
XIA
2015. All rights reserved.
xviii
The L
IST
M
ODE
T
RACES
display shows the data from the binary list mode files (.bin). If
waveforms were collected, they are shown in the graph section of the panel. Event and channel
header information – energy, time stamps, and hit patterns as described in section 4.1.2 – are
shown in the fields above the graph section. After specifying a data file with the
Find
button,
you can select an event to view by entering its number in the
Event Number
field.
The button
Digital Filters
opens a new plot that shows the response of the trigger filter and
energy filter computed from the list mode waveforms. This plot is more precise than the related
graph opened from the O
SCILLOSCOPE
since it uses the same full rate (13.3ns) data as the filters
implemented in the module, not the reduced rate sampled at the
OSCILLOSCOPE
’
S
dT
. However,
unless long list mode traces are acquired or energy filters are short, there may not be sufficient
data to compute the energy filter properly.
The L
IST
M
ODE
S
PECTRUM
display is a plot similar to the MCA
S
PECTRUM
, but it is computed
from the energies saved in the list mode data file. Since energies are stored there in full 16 bit
precision, binning can be made finer than in the MCA
S
PECTRUM
, which is limited to 32K bins.
See the online help for a detailed description of the controls.
3.4.3 Run Statistics
Figure 3.8: The R
UN
S
TATISTICS
panel.
The R
UN
S
TATISTICS
panel shows the live times and count rates measured by the Pixie-4. The
numbers can be updated by clicking the
Update
button and read from or save to
Files
. For a
detailed description of the definition of these values, see section 6.6.
3.4.4 File Series
See section 3.6 for a more detailed description