Multi-frame acquisition with frame trigger
Multi-frame acquisition with burst frame trigger
AcquisitionMode = MultiFrame;
AcquisitionFrameCount = 2;
TriggerMode = ExternalTriggered;
TriggerInput = Line1;
AcquisitionBurstFrameCount = 1;
AcquisitionStart();
Frame 1
Frame 2
AcquisitionStart();
Line1 trigger 1
Line1 trigger 2
Frame 3
Frame 4
2 frames are acquired.
A trigger is necessary
for each of them.
AcquisitionStop() is
optional.
Line1 trigger 3
Line1 trigger 4
AcquisitionMode = MultiFrame;
AcquisitionFrameCount = 3;
TriggerMode = ExternalTriggered;
TriggerInput = Line1;
AcquisitionBurstFrameCount = 2;
AcquisitionStart();
Frame 1
Frame 2
Line1 trigger 1
Frame 3
For each trigger, 2
frames are acquired.
Line1 trigger 2
AcquisitionFrameCount
frames are acquired.
Acquisition is stopped