Operation
Over Voltage Protection (OVP)
Release 2.2
41
Remote
Programming
of OVP With
External
Voltage
Sources
To remotely program the OVP trip level using a 0-5V DC (
Figure 3.2
) or 0-10V DC
(
Figure 3.3
) voltage source use the following procedure.
1. With the unit off and disconnected from its AC source, remove the cover and set
switch SW1-4 closed (default factory setting) for 0-5V programming or open for
0-10V programming. Also check that switch SW1-8 is closed (default factory
setting). Set the front panel OVP adjusting potentiometer fully clockwise (until
audible clicking is heard or 20 turns maximum).
2. Remove the default jumper connecting pins 16 and 3 of connector J3 and connect
the voltage source between pins 3 (positive) and 12 (negative). Set the
programming source voltage to maximum.
3. Turn the unit on and adjust the output to the desired trip voltage.
4. Slowly reduce the programming voltage until the red OVP indicator lamp lights
and the power supply shuts down.
5. Turn the POWER switch to OFF.
6. Turn the voltage control knob to minimum.
7. Turn the POWER switch back ON and increase the voltage to check that the
power supply shuts off the output at the desired voltage.
Figure 3.2
SW1/J3 Configuration for 0-5Vdc OVP Programming
(J3 sense line, voltage and current control jumpers shown set for local operation)
Содержание XKW 150-7
Страница 2: ......
Страница 3: ...Operating Manual for XKW 1000 Watt Series Programmable DC Power Supply ...
Страница 15: ...Features and Specifications Front Panel Controls Release 2 2 13 Figure 1 1 Power Supply Front Panel ...
Страница 24: ...Features and Specifications Specifications 22 Operating Manual for XKW 1kW Series Power Supply ...
Страница 38: ...Installation Local and Remote Sensing 36 Operating Manual for XKW 1kW Series Power Supply ...
Страница 56: ...Operation Using Over Temperature Protection OTP 54 Operating Manual for XKW 1kW Series Power Supply ...
Страница 59: ......