Scan Measurement
To prepare your
i1
Pro device for a scan measurement you need to mount the i1Pro
device on the ruler. Make sure your device is calibrated and the
status indicators
are pulsating white before you mount it.
1. Unfold the backup board.
2. Use the clamp to secure the test chart on the backup board.
3. Place the ruler on the backup board.
4. Position the
i1
Pro device in the carriage on the ruler. Make sure that the positioning
pins on the carriage slide into the reference gaps on the bottom of the
i1
Pro device.
Double-check that the
i1
Pro device seats firmly in the carriage.
5. During scanning measurement make sure that the device’s limiting aperture glides
smoothly on the guide rail of the ruler.
The distance between the
i1
Pro device and the surface of the printing substrate
is critical for the accuracy of your measurements. If the limiting aperture of the
device or the edge of the ruler‘s guide rail is worn out you should replace them.
Spare parts can be ordered through X-Rite‘s Service Centers.