18
Alternating current measurement
▸
Set the selector switch
[9]
to
V~
.
▸
Plug the black measuring lead into the COM jack
[5]
.
▸
Plug the red measuring lead into the V/Ω input
jack
[6]
.
▸
Connect the measuring leads to the test object.
3
The multimeter automatically searches for the most
favourable measurement range.
3
The measurement result is output at the display
[1]
.
AC voltage
Measuring
range
Resolution
Accuracy
400 mV
0.1 mV
± 1.5 % rdg + 6 d
400 mV
0.1 mV
± 0.8 % rdg + 8 d
4 V
1 mV
40 V
10 mV
50 V
100 mV
Input resistance: 10 MΩ.
Overvoltage protection: 120 V
RMS
Direct current measurement
▸
Set the selector switch
[9]
measuring range to
mA
or
A
.
▸
Plug the black measuring lead into the COM jack
[5]
.
▸
Plug the red measuring lead into the mA meas-
uring jack
[8]
or the 20 A jack
[7]
.
Attention !
For a current above 400 mA, the 20 A measuring
jack
[7]
must be used!
▸
Connect the measuring leads to the test object.
3
The multimeter automatically searches for the most
favourable measurement range.
3
The measurement result is output at the display
[1]
. The current direction is indicated by the lead-
ing sign.
Direct current
Measuring
range
Resolution
Accuracy
40 μA
10 μA
± 1.0 % rdg + 5 d
400 μA
100 μA
4 A
1 mA
± 1.2 % rdg + 10 d
20
*
A
10 mA
Attention !
*
To protect the device against overheating, allow
a pause of 15 minutes for cooling after measuring
for a period of no more than 10 seconds.
Overload protection:
•
The mA range is protected by a 400 mA self-
resetting fuse.
•
Observe the maximum duty cycle in the 20 A
range. Fuse: F 20 A
Alternating current measurement
▸
Set the selector switch
[9]
measuring range to
“
mA
“ or “
A
”.
▸
Plug the black measuring lead into the COM jack
[5]
.
▸
Plug the red measuring lead into the mA meas-
uring jack
[8]
or the 20 A jack
[7]
.
Attention !
For a current above 400 mA, the 20 A measuring
jack
[7]
must be used!
▸
Connect the measuring leads to the test object.
3
The multimeter automatically searches for the most
favourable measurement range.
3
The measurement result is output at the display
[1]
.
Attention !
T
o protect the device against overheating, allow a
pause of 15 minutes for cooling after measuring
for a period of no more than 10 seconds.
Alternating current
Measuring
range
Resolution
Accuracy
40 mA
10 μA
± 1.2 % rdg + 8 d
400 mA
100 μA
4 mA
1 mA
± 2.0 % rdg + 5 d
20
*
A
10 mA
Attention !
T
o protect the device against overheating, allow a
pause of 15 minutes for cooling after measuring
for a period of no more than 10 seconds.
Overload protection:
•
The mA range is protected by a 400 mA self-
resetting fuse.
•
Observe the maximum duty cycle in the 20 A
range. Fuse: F 20 A
Содержание 0715 53 440
Страница 3: ...1 2 3 4 5 6 7 9 8 12 11 10 I 3...
Страница 4: ...4...
Страница 104: ...104 GR W rth 3 1...
Страница 105: ...105 50 V AC 120 V DC 30 IR 50 V AC 120 V DC W rth...
Страница 176: ...176 BG W rth 3 1...
Страница 177: ...177 50 V AC 120 V DC 30 IR 50 V AC 120 V DC W rth...
Страница 212: ...212 RU W rth 3 1...
Страница 213: ...213 50 120 30 50 120 W rth...
Страница 219: ...219 3 1 36 PULSE 9 PULSE CYL 4 COM 5 6 3 1 PULSE 0 1 40 0 01 2 0 5 120 HOLD HOLD HOLD 3 3 3 AUTO RAN 11 RAN 1 9 6F22...
Страница 239: ......