Appendix 2:
Turbidity values under 1 FNU/NTU
With turbidity values (under 1 FNU/NTU), the measured value is
strongly influenced by the cell and its alignment.
In order to increase measuring accuracy with turbidity values under 1
FNU/NTU, calibration in the 0.02 FNU/NTU standard and later
measurement should be carried out in the same cell. For calibration in
the standards 10.0 and 1000 FNU/NTU follow the instructions on the
display.
Proceed as follows to measure turbidity values under 1 FU/NTU:
Calibration procedure:
Calibrate
after the calibration interval has expired
with a temperature change
Measuring
1
Press the
<CAL/ZERO>
key.
The menu-guided calibration begins.
2
Carry out calibration of the standards 1000 FNU/NTU and
10.0 FNU/NTU in a clean and unscratched cell according to
the menu guidance.
3
Fill the cleaned cell with the 0.02 FNU/NTU standard and
calibrate.
4
Mark the alignment of the cell.
5
Fill the marked and cleaned cell with test sample, align it
with the marking and measure.
6
If necessary, fill the marked and cleaned cell once more
with test sample and carry out further measurements.
Содержание Turb 430 IR/T
Страница 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Страница 2: ...ba75507e05 07 2016 Turb 430 IR T Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Страница 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Страница 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Страница 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Страница 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Страница 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Страница 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Страница 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...
Страница 71: ...Turb 430 IR T ba75507e05 07 2016 71...