Turb 430 IR/T
Technical data
59
ba75507e05
07/2016
7.2
Turbidity
7.2.1
Turb 430 IR
7.2.2
Turb 430 T
Measuring principle
Nephelometric measurement according to DIN EN ISO 7027
Light source
Infrared LED
Measuring range
0.01 ... 1100 FNU/NTU
Resolution
in the range
0,01 ... 9,99
max. 0.01 FNU/NTU
in the range
10,0 ... 99,9
max. 0.1 FNU/NTU
in the range
100 ... 1100
max. 1 FNU/NTU
Accuracy
in the range
0 ... 1100 FNU/NTU
± 2 % of the measured value
or ± 0.01 FNU/NTU
Reproducibility
0,5% of the measured value
Response time
4 seconds
Calibration
Automatic 3-point calibration
Measuring principle
Nephelometric measurement according to US EPA 180.1
Light source
White light tungsten lamp
Measuring range
0.01 ... 1100 NTU
Resolution
in the range
0,01 ... 9,99
max. 0.01 NTU
in the range
10,0 ... 99,9
max. 0.1 NTU
in the range
100 ... 1100
max. 1 NTU
Accuracy
in the range
0 ... 500 NTU
± 2 % of the measured value
or ± 0.01 NTU
in the range
500 ... 1100 NTU
± 3 % of the measured value
Reproducibility
1% of the measured value
Response time
7 seconds
Calibration
Automatic 3-point calibration
Содержание Turb 430 IR/T
Страница 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Страница 2: ...ba75507e05 07 2016 Turb 430 IR T Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Страница 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Страница 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Страница 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Страница 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Страница 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Страница 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Страница 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...
Страница 71: ...Turb 430 IR T ba75507e05 07 2016 71...