4.6
Saving
The meter has 1000 memory locations for measurement datsets.
You can transmit measured values (datasets) to the data memory with
the
<STO>
key.
Each storage process transmits the current dataset to the interface at
the same time.
The number of memory locations that are still free is displayed in the
Store
menu. The number of memory locations that are occupied is
displayed in the
System
Measured value memory
menu.
Measurement dataset
A complete dataset consists of:
Date/time
ID number (ID)
Measured value
4.6.1
Storing measurement datsets
Proceed as follows to transmit to the data memory and simultaneously
output to the interface a measurement dataset:
Note
A measurement dataset is stored quickly by twice pressing
<STO>
. It
is stored with the ID last set.
1
Press the
<STO>
key.
The
Store
display appears.
2
Using
<
?
> <
?
>
,
<START/ENTER>
and the number keys,
change and confirm the ID number (
ID
) as necessary (0 ...
999).
3
Using
<START/ENTER>
or
<STO>
, confirm
Store
.
The dataset is stored. The instrument switches to the
measured value display.
i
02.02.2005 11:24:16
16.80 FNU/NTU
ID
:
1
Store
(ID: 1)
Store
(
996
free
)
Содержание Turb 430 IR/T
Страница 1: ...OPERATING MANUAL ba75507e05 07 2016 Turb 430 IR T HANDHELD TURBIDIMETER...
Страница 2: ...ba75507e05 07 2016 Turb 430 IR T Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 12: ...Safety Turb 430 IR T 12 ba75507e05 07 2016...
Страница 16: ...Commissioning Turb 430 IR T 16 ba75507e05 07 2016 LabStation...
Страница 18: ...Commissioning Turb 430 IR T 18 ba75507e05 07 2016...
Страница 50: ...Operation Turb 430 IR T 50 ba75507e04 06 2009...
Страница 56: ...What to do if Turb 430 IR T 56 ba75507e05 07 2016...
Страница 60: ...Technical data Turb 430 IR T 60 ba75507e05 07 2016...
Страница 66: ...Lists Turb 430 IR T 66 ba75507e05 07 2016...
Страница 68: ...Index Turb 430 IR T 68 ba75507e05 07 2016...
Страница 71: ...Turb 430 IR T ba75507e05 07 2016 71...