Commissioning
photoLab
®
7100 VIS
12
ba77095e03
10/2017
3
Commissioning
3.1
General notes on handling
The photoLab
®
7100 VIS
photometer is an optical precision meter. There-
fore, it should always be handled with care, especially in mobile use. Always
protect the meter from conditions that could damage the mechanical, optical
and electronic components. Heed the following points especially:
The temperature and humidity during operation and storage must be
within the limits specified in chapter 7 T
ECHNICAL
DATA
.
The following influences always have to be avoided with the meter:
– Extreme dust, moisture and wetness
– Exposure to intensive light and heat
– Fumes that are corrosive or contain high concentrations of solvents.
For measuring, the meter must be placed on a flat surface.
Spilled liquid or other material should be removed immediately (see sec-
tion 5.3 C
LEANING
).
If a cell has broken in the cell shaft, the cell shaft should be cleaned imme-
diately (see section 6.1 A
CTIONS
IN
THE
CASE
OF
A
BROKEN
CELL
).
The cell shaft should always be closed when the photometer is not used.
For the instrument to be transported the cell shaft has to be empty.
For mobile used we recommend using the suitable transport case
FC spectral 6000 .
3.2
Initial commissioning
Perform the following activities:
Insert the buffer batteries (see section 3.2.1)
Connect the power supply (see section 3.2.2)
Switch on the photometer (see section 3.2.3)
Set the language (see section 3.2.3)
Set the date and time (see section 3.2.4)
Carry out a zero adjustment (see section 4.3)
Packing
This photometer is sent out in a protective transport packing.
CAUTION
Keep the original packing including the inner packing to protect the in-
strument against hard shocks if it has to be transported.