photoLab
®
7100 VIS
What to do if...
39
ba77095e03
10/2017
Obviously incorrect
measured values
Fluctuating
measured values
Self test failed.
– The cell shaft is contaminated
– Clean the cell shaft (see section
5.3.2 and section 6.1)
– Restart the instrument
– If necessary, confirm the
Please
remove cell
message with
<START·ENTER>
.
– Instrument defective
– Please contact the service
department.
Cause
Remedy
Cause
Remedy
– Cell contaminated
– Clean the cell
– Dilution set incorrectly
– Set the dilution
– Selected method not suitable
– Select different method
– Zero measurement incorrect
– Perform zero measurement
– Blank value incorrect
– Remeasure the blank value
– Lamp module exchanged
– Reset the service hour counter
for the lamp module.
Cause
Remedy
– Cell shaft cover open
– Close the cell shaft cover
– Lamp replacement
– Reset the service hour counter
for the lamp module.
Cause
Remedy
–
System test
: Instrument defective
– Please contact the service
department.
–
Filter test
: Instrument defective
– Please contact the service
department.