photoLab
®
7100 VIS
Maintenance and cleaning
33
ba77095e03
10/2017
Note
After recommissioning, carry out a new zero adjustment for all measure-
ments.
3
Disconnect the electrical plug con-
nection (3). Disconnecting it may
take a lot of effort.
4
Unscrew the two knurled-head
screws (4) and remove the defec-
tive lamp module (5).
5
Fix the new lamp module with the
knurled-head screws. When doing
so, the metal-plated side of the
PCB must point outward, i. e.
toward the knurled-head screws.
6
Reconnect the electrical plug con-
nection.
7
Fix the lamp cover.
8
Reset the service hour counter for
the lamp module.
3
4
5