Configuration
109
Note:
The recorder range 2 (Rec 2) can be selected regardless of the setting for channel A in the
difference operation mode as well:
In this case the adjusted range is placed symmetrically around the zero point, which is at 10
mA for 0-20 mA and at 12 mA for 4-20 mA. In the difference operation mode D the current
output of Rec 2 is, therefore, at 10 mA or 12 mA respectively with similar pH or mV values of
channel A and B (difference A-B).
Example: Difference operation D: Signal output at recorder outputs REC 1 and REC 2
for output current 0-20 mA or 4-20 mA
Settings:
Recorder measuring range REC 1:2.00 ... 12.00 pH
Recorder measuring range REC 2:
±
5 pH
Recorder output Rec 2 depending on test signal in difference operation mode:
Output current 0-20 mA
Output current 4-20 mA
Measured values
Recorder outputs
Measured values
Recorder outputs
Chan. A
Chan. B Difference
REC 1
REC 2
Chan. A
Chan. B Difference
REC 1
REC 2
7.00 pH
7.00 pH
0.00 pH 10.00 mA 10.00 mA
7.00 pH
7.00 pH
0.00 pH 12.00 mA 12.00 mA
5.00 pH
7.00 pH
-2.00 pH
6.00 mA
6.00 mA
5.00 pH
7.00 pH
-2.00 pH
8.80 mA
8.80 mA
7.00 pH
5.00 pH
2.00 pH 10.00 mA 14.00 mA
7.00 pH
5.00 pH
2.00 pH 12.00 mA 15.20 mA
2.00 pH
6.00 pH
-4.00 pH
0.00 mA
2.00 mA
2.00 pH
6.00 pH
-4.00 pH
4.00 mA
5.60 mA
6.00 pH
2.00 pH
4.00 pH
8.00 mA 18.00 mA
6.00 pH
2.00 pH
4.00 pH 10.40 mA 18.40 mA
Relay functions: Selection as transmitting contacts
(R models)
A function can be assigned to each relay (see following chart).
Setting:
Submenu "CONF RELAYS“.
With
enter submenu
CONF RELAYS.
Содержание pH 171 2K
Страница 2: ...2 ...
Страница 68: ...Technische Daten 68 EMV Zertifikat ...
Страница 76: ...76 ...
Страница 82: ...86 ...
Страница 148: ...Technical Data 152 EMC certificate ...
Страница 152: ...Connections of the different models 156 1 channel instruments ST model standard R model RS model R RS model ...
Страница 153: ...Connections of the different models 157 2 channel instruments 2K model 2K R model 2K RS model 2K R RS model ...
Страница 156: ...160 ...