Calibration
129
If the temperature compensation has been configured to manual input (Configuration: T-C
MANUAL) the following is displayed:
With
set temperature of buffer
solution 1.
(Default value 15°C).
With
accept and start calibration.
Measurement runs until a stable value is
achieved.
Length of the calibration process:
min. 30 s up to max. 10 minutes.
(If after max. 10 minutes no stable value is
achieved the error message "INSTABLE"
appears, see chapter "Indication after invalid
calibration").
Stable measured value for buffer solution 1
accepted.
Rinse electrode and immerse into buffer
solution 2.
With
start calibration with
buffer solution 2.
If the temperature compensation has been configured to manual input (Configuration: T-C
MANUAL) the following is displayed:
With
set temperature of buffer
solution 2.
(Default value 15°C).
With
accept and start calibration with
buffer solution 2.