Configuration
108
Setting of recorder output Rec 2:
Submenu "CONF REC 2".
With
enter submenu
CONF REC 2.
Submenu "CONF REC 2":
Upper rating, e. g. 12.00 pH is flashing.
With
set upper limit of measuring range
between 4.00 and 14.00 pH.
With
accept rating and go on to lower limit
of measuring range.
Upper rating, e. g. 2.00 pH is flashing.
With
set lower limit of measuring range
between 0.00 and 10.00 pH.
With
accept indicated value and go on to
CONF RELAYS.
Indication for parallel operation (e. g. Redox measurement)
With
set upper limit of measuring range
between -500 mV and +1000 mV.
With
accept indicated value.
With
set lower limit of measuring range
between -1000 mV and +500 mV.
With
accept indicated value and go on to
CONF RELAYS.
Indication for difference operation
With
select desired difference range for
0/4-20mA:
With
accept indicated values and go on to
submenu CONF RELAYS.