Parameterization
121
Parameterization of timer function
(R models)
In the PARA TIMER submenu the times T 1 and T 2 have to be set for the timer function (see also
chapter "Configuration: Selection as transmitting relays" CONF RELAYS)
Note:
This menu only appears if a relay - relay 1 in the following example - has been
configured as a timer (Configuration level: submenu "CONF RELAYS").
Setting:
Submenu "PARA TIMER".
With
enter submenu
"PARA TIMER".
With
set desired time T1.
Setting range: 1...24 hours.
With
accept and go on to "T2/min".
⇐
relay number 1.
With
set desired time T2.
Setting range:
0.5 ... 5.0 minutes,
Step size:
0.5 minutes.
With
accept and go on to next submenu
"PARA RS 485" (RS models) or "PARA PASS"
or to setting of relay no. 2.
Parameterization for further relays configured as timers is made as described above.