Settings
Oxi 3315
34
ba75944e02
03/2012
6.2
Sensor-independent settings
6.2.1
System
To open the
Storage & config
menu, press the
<ENTER_>
key in the measured
value display. After completing the settings, switch to the measured value dis-
play with
<M>
.
Default settings are printed in
bold
.
Response time
t90
30
... 300 s
Response time of the signal filter (in
seconds).
A signal filter in the sensor reduces the
limits of variation of the measured
value. The signal filter is characterized
by the response time t90. This is the
time after which 90 % of a signal
change is displayed.
Reset
-
Resets all sensor settings to the deliv-
ery condition (see section 6.3.1 R
36).
Menu item
Possible
setting
Description
Menu item
Possible setting
Description
System
/
General
/
Lan-
guage
Deutsch
English
(further)
Selects the menu language
System
/
General
/
Acous-
tic signal
on
off
Switches on/off the beep on keystroke
System
/
General
/
Illumi-
nation
Auto
on
off
Switches the display illumination on/off
System
/
General
/
Con-
trast
0 ...
50
... 100
Changing the display contrast
System
/
General
/
Switchoff time
10 min ...
1h
...
24 h
Adjusts the switch-off time
System
/
General
/
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperatures are displayed with the selected
unit.
System
/
General
/
Stability control
on
off
Switches on or off the automatic stability control
during measurement (see section 6.2.3 A