Oxi 3315
Dissolved oxygen
ba75944e02
03/2012
31
5.3.6
Calibration data
Displaying the
cali-
bration record
The calibration record of the last calibration is to be found under the menu item,
Calibration
/
Calibration record
. To open it in the measured value display, press
the
<CAL_>
key.
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration
/
Calibration data storage
/
Display
. To open the
Calibration
menu,
press the
<ENTER>
key in the measured value display.
Calibration evalua-
tion
After the calibration, the meter automatically evaluates the current status of the
6.
Using
<
> <
>
, set the correction factor to adjust the displayed con-
centration value to the nominal value (value of the comparison mea-
surement). Subsequently, accept the correction factor with
<ENTER>
.
The measuring instrument switches to the measured value display.
The status display
[Factor]
is active.
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/
function
Description
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
Output all calibration records to
the interface with
<F2__>
[USB
output]
.
Quit the display with
<F1>
/
[Zurück] or
<ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to RS232/
USB
-
Outputs the calibration data stor-
age to the interface