Operation
Oxi 3315
22
ba75944e02
03/2012
4.4.5
Example 2 on navigation: Setting the date and time
The measuring instrument has a clock with a date function. The date and time
are indicated in the status line of the measured value display.
When storing measured values and calibrating, the current date and time are
automatically stored as well.
The correct setting of the date and time and date format is important for the fol-
lowing functions and displays:
Current date and time
Calibration date
Check date (for IDS D.O. sensors)
Identification of stored measured values.
Therefore, check the time at regular intervals.
Setting the date,
time and date format
The date format can be switched from the display of day, month, year
(
dd.mm.yy
) to the display of month, day, year (
mm/dd/yy
or
mm.dd.yy
).
After a fall of the supply voltage (empty batteries), the date and time
are reset to 01.05.2011 00, 00:00 hours.
1.
In the measured value display:
Open the
Storage & config
menu with
<ENTER_>
.
The instrument is in the setting mode.
2.
Select and confirm the
System
/
Clock function
menu with
<
><
>
and
<ENTER>
.
The setting menu for the date and time opens up.
3.
Select and confirm the
Time
menu with
<
><
>
and
<ENTER>
.
The hours are highlighted.
4.
Change and confirm the setting with
<
><
>
and
<ENTER>
.
The minutes are highlighted.
5.
Change and confirm the setting with
<
><
>
and
<ENTER>
.
The seconds are highlighted.
Date format:
dd.mm.yy
Date:
01.09.2011
Time:
14:53:40
Clock function
01.09.2011
08:00
Back