Oxi 3315
Data storage
ba75944e02
03/2012
39
7.2
Automatic storage at intervals
The storage interval (
Interval
) determines the time interval between automatic
storage processes. Each storage process transmits the current dataset to the
interface at the same time.
Configuring the
automatic storage
function
Settings
You can configure the automatic storage function with the following settings:
Starting the auto-
matic storage func-
tion
To start the automatic data storage function, select
continue
with
<
><
>
and
1.
Press the
<STO_>
key.
The menu for automatic storage appears.
1 Adjusted entire storage duration
2 Max. available storage duration
3 Graphical display of the storage usage
ID number
1
Interval 30
s
Duration 180
min
continue
0d03h00min
0
1d17h33min
Automatic data storage
1
2
3
01.09.2011
08:00
Back
Menu item
Possible setting
Description
ID number
1 ... 10000
ID number for the dataset series.
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min, 30 min,
60 min
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the stor-
age duration.
Duration
1 min ... x min
Storage duration.
Specifies after which time the auto-
matic storage should be terminated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.