Data storage
Oxi 3315
38
ba75944e02
03/2012
7
Data storage
You can transmit measured values (datasets) to the data storage:
Manual data storage (see section 7.1 M
Automatic data storage at intervals (see section 7.2 A
Each storage process transmits the current dataset to the interface at the same
time.
7.1
Manual storage
You can transmit a measurement dataset to the data storage as follows. The
dataset is at the same time output to the interface:
If the storage is full
When all storage locations are occupied, it is not possible to continue storing.
Then you can e.g. transmit the stored data to a PC (see section 7.3.1 E
, page 41) and subsequently erase the data
storage (see section 7.3.2 E
42).
1.
Press the
<STO>
key shortly.
The menu for manual storage appears.
2.
If necessary, change and confirm the ID number (1 ... 10000) with
<
><
>
and
<ENTER>
.
The dataset is stored. The meter switches to the measured value dis-
play.
01.09.2011 07:43:33
Ox 7.88 mg/l 24.8 °C AR +++
t90 = 30 s
ID number:
1
continue
Manual data storage 4 von 500
01.09.2011
08:00
Back