Turbidity measurement (VisoTurb
®
900-P)
inoLab
®
Multi 9310 IDS
62
ba75937e10
08/2021
9
Turbidity measurement (VisoTurb
®
900-P)
9.1
Measuring
9.1.1 Measuring the turbidity
Preparatory activi-
ties
Perform the following preparatory activities when you want to measure:
Avoid gas bubbles (e.g. air bubbles) in the test sample.
Use suitable vessels for measurement and calibration
(see operating manual of the VisoTurb
®
900-P sensor).
Heed the minimum depth of immersion for the sensor
Measuring
You can carry out turbidity measurements as follows:
The sensor connection and the USB-B (device) interface are gal-
vanically insulated. This facilitates interference-free measurements
also in the following cases:
Measurement in grounded test samples
Measurement with several sensors connected to one
Multi 9310 IDS in one test sample
1.
Connect a turbidity sensor to the measuring instrument.
The turbidity measuring screen is displayed.
The data for the connected IDS turbidity sensor are automatically taken
over.
2.
Fill the test sample into a lightproof measuring beaker up to a level of
at least 6 cm.
3.
When immersing the sensor in the test sample, hold the sensor at an
angle.
4.
For measuring, position the sensor upright.
5.
Position the sensor in a way that meets the following requirements.
Distance to the bottom: 6 cm
Distance to the walls of the beaker: 2 cm
Minimum depth of immersion: 2 cm
To position the sensor optimally and permanently while it is mea-
suring, fix it on a support.
1.
Perform the preparatory activities.
2.
Immerse the sensor in the test sample at an angle and then position it
in the measuring container.