inoLab
®
Multi 9310 IDS
Conductivity
ba75937e10
08/2021
61
Calibration evalua-
tion
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Calibration record
(
USB output
)
Menu item
Setting/
function
Explanation
Calibration
/
Calibration data
storage
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibration
record to the interface with
<F2>
/
[USB output]
.
Output all calibration records to
the interface with
<F2__>
[USB
output]
.
Quit the display with
<F1>
/
[Back] or
<ENTER>
.
Switch directly to the measured
value display with
<M>
.
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the calibration data storage
to the interface (see section 12
T
Display
Calibration record
Cell constant [cm
-1
]
+++
Within the range
0.450 ... 0.500 cm
-1
Error
Error
Outside the range
0.450 ... 0.500 cm
-1
Error elimination (see section 15
W
CALIBRATION Cond
03.04.2013 07:43:33
TetraCon 925
Ser. no. 09250033
Cell constant 0.476 1/cm 25.0 °C
Sensor
+++