DIQ/S 181
Outputs
85
ba77083e01
02/2015
6.5
Setting tables for current outputs
6.5.1
Functions and settings
To set a current output, its function has to be selected first. Then the
relevant setting table is displayed:
Function of current
outputs and relevant
setting tables
6.5.2
Recorder
Function
The measured values of the linked sensor at the current output are set
up as current intensity in the
Recorder
application. The output of
measured values is laid down in the
Recorder type
,
Start value
and
End
value
settings.
Settings
Function
Setting table
No function
The current output is not used.
Recorder
Fixed current value
Setting
Selection/Values
Explanation
Recorder type
0 to 20 mA or
4 to 20 mA
Start value
(sensor
dependent)
Minimum spacing:
≤
5 % of the measuring
range
(sensor-dependent)
End value
Measured variable
Main variable
Adjoining variable
Main variable
designates
the actual measured
parameter of the sensor
(e.g. pH, oxygen, etc.).
Adjoining variable
designates an additional
measured parameter
(e.g. temperature).
Attenuation
0 ... 40 mA/s
Speed of change of the
output current (mA/s) in
the case of erratic
changes of the input sig-
nal.
Содержание DIQ/S 181
Страница 1: ...OPERATING MANUAL ba77083e01 02 2015 DIQ S 181 UNIVERSAL TRANSMITTER FOR ONE DIGITAL SENSOR...
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...List of contents DIQ S 181 6 ba77083e01 02 2015...
Страница 102: ...Index DIQ S 181 102 ba77083e01 02 2015...
Страница 103: ......