Behavior of linked outputs
6.6.1
Behavior in case of error
For linked relay outputs or current outputs, you can determine the
behavior of the outputs in case of errors. Depending on the use of the
output, the behavior in case of errors is set in the following menus:
The specified behavior occurs with the following errors.
The display shows
Init
,
----
,
Error
or
OFL
.
The meaning of these displays is given in section 4.2 of this operat-
ing manual.
The supply voltage is too low.
6.6.2
Behavior in non-operative condition
An output is non-operative when no function is activated for the output.
An output becomes non-operative in case of
Power failure
(As soon as the supply voltage is sufficient again, the non-operative
condition of the outputs ends. The outputs function as specified by
the user again.)
Erasing the link with the sensor
Changing the
Measuring mode
sensor setting for a linked sensor
Changing the
Measuring range
sensor setting for a linked sensor
Changing the
Temperature mode
sensor setting for a linked sensor
Settings in the non-
active condition
Output
Menu
Limit indicator
Behavior at error
Frequency controller
Frequency with error
Pulse-width output
Error pulse width
(see section 6.4.6)
Recorder
Current with error
Before editing sensor settings a note appears on the
display to inform you that links will be erased when you
change the
Measuring mode
or
Measuring range
sensor
setting.
Current output
Relay output
Current: 0 A
Relay: Open
Содержание DIQ/S 181
Страница 1: ...OPERATING MANUAL ba77083e01 02 2015 DIQ S 181 UNIVERSAL TRANSMITTER FOR ONE DIGITAL SENSOR...
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...List of contents DIQ S 181 6 ba77083e01 02 2015...
Страница 102: ...Index DIQ S 181 102 ba77083e01 02 2015...
Страница 103: ......