8.1
Information on errors
Log book
The universal transmitter DIQ/S 181 performs a comprehensive cycli-
cal self test during operation. When doing so, the DIQ/S 181 identifies
all states that deviate from normal operation and enters corresponding
messages in the log book (information or error message).
With the aid of the log book, you can call up instructions on how to clear
the error directly on the terminal. The log book is described in detail in
section 4.4.3 L
8.2
Error causes and remedies
The DIQ/S 181 no longer
reacts to entries
"Error" in the measured
value display
The selected system
language was not
activated for a sensor
Information on possible errors of the sensor is given in
chapter W
HAT
TO
DO
IF
... of the sensor operating manual.
Cause
Remedy
– System error
Reset the DIQ/S 181:
– Switch off the power supply
and switch it on again after
10 s
Cause
Remedy
– Communication with the sensor
interrupted
– Check the cable connection
– Error in the sensor
– Switch off the power supply
and switch it on again after
10 s
Cause
Remedy
– A system language was selected
that is not available in the sen-
sor.
The English language was acti-
vated instead of the selected
system language.
– Contact WTW,
a software update is
required for the respective
components
Содержание DIQ/S 181
Страница 1: ...OPERATING MANUAL ba77083e01 02 2015 DIQ S 181 UNIVERSAL TRANSMITTER FOR ONE DIGITAL SENSOR...
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...List of contents DIQ S 181 6 ba77083e01 02 2015...
Страница 102: ...Index DIQ S 181 102 ba77083e01 02 2015...
Страница 103: ......