background image

ISO-HPO-2

World Precision Instruments 

3

Gas permeable membrane

–The membrane covers the end of the stainless sleeve 

to separate it from the external environment.

Stainless steel sleeve

 (WPI#

600012, 

pkg. of 4)–The sleeve houses the sensitive 

electrode pair. The sleeve is flanged to properly connect with the locking cap. It must 

contain fresh filling solution (WPI#

100042

).

Working/Counter electrode

–This is an internal H

2

O

2

-sensing pair of working and 

counter (reference) electrodes.

Locking cap–

This cap attaches the sleeve to the probe handle.

When the sensor is assembled (with locking cap and sleeve in place) the internal 

electrode should press gently against the polymeric membrane, which will be 

slightly 

stretched

. This ensures that the electrolyte diffusion is as thin as possible, minimizing 

sensor response time. 

NOTE

: Once a membrane is stretched it is permanently deformed and cannot be 

reused if the sleeve is removed from the electrode. 
Additional membrane sleeves are available in packages of 4 (WPI #

600012

). The 

ISO-HPO-2

 start-up kit (WPI #

600011

) also includes replacement membrane sleeves, 

along with all the accessories to fill them properly with filling solution (electrolyte). 

OPERATING INSTRUCTIONS

Polarizing the Sensor

1.  Measure an appropriate volume (for example, 10mL) of 0.1M PBS buffer solution 

into a glass vial. Place the vial on a magnetic stirring plate and put a small stir bar 

into the vial. Set the stir rate at a medium speed. 

2.   Remove the sensor from the filling solution in which the tip has been immersed 

during storage. See the 

Probe Unpacking Instructions

 included with your sensor. 

 

NOTE

: Some sensors are shipped or stored dry. For these sensors, you will need 

to install the membrane sleeve (included). See “Replacing the Membrane Sleeve” 

on page 7.

3  Immerse the 

ISO-HPO-2

 sensor tip in PBS buffer solution. The sensor tip should 

be immersed about 0.3-0.5mm into the solution, and should not be touched by 

the stir bar.

4.  Connect the 

ISO-HPO-2

 sensor to the input on the free radical analyzer.

5.  Turn on the free radical analyzer (

TBR4100/1025)

6.  Select the poise voltage for hydrogen peroxide on the free radical analyzer. On 

the 

TBR4100/1025

, set the 

Probe Select

 switch to 

HPO

 (+450mV). 

 

NOTE

: It is important that the poise be set correctly to avoid erroneous results. 

7.  Set the current range to 10nA

Содержание ISO-HPO-2

Страница 1: ...WORLD PRECISION INSTRUMENTS ISO HPO 2 Hydrogen Peroxide Sensor for use with WPI Analyzers Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 031919...

Страница 2: ......

Страница 3: ...king of 0 1M PBS Buffer Solution 5 SENSOR CARE AND MAINTENANCE 6 Durability and Handling 6 Storing the Sensor 7 Cleaning the Membrane 7 Replacing the Membrane Sleeve 7 ACCESSORIES 9 TROUBLESHOOTING 10...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...working reference electrode combination This electrode fits inside a disposable protective stainless steel sleeve WPI 600012 which must contain fresh electrolyte WPI 100042 and is separated from the...

Страница 6: ...if any parts are missing at 941 371 1003 or customerservice wpiinc com Returns Do not return any goods to WPI without obtaining prior approval RMA required and instructions from WPI s Returns Departme...

Страница 7: ...includes replacement membrane sleeves along with all the accessories to fill them properly with filling solution electrolyte OPERATING INSTRUCTIONS Polarizing the Sensor 1 Measure an appropriate volum...

Страница 8: ...f 450mV The oxidation of H2 O2 at the sensor surface produces a small current pA which is detected by the TBR4100 1025 The amount of current produced is linearly proportional to amount of H2 O2 in the...

Страница 9: ...er a half hour in solution it is likely that the membrane has been damaged and the sleeve needs to be changed See Replacing the Mem brane Sleeve on page 7 NOTE The calibration should be carried out at...

Страница 10: ...nsitivity However if the pH is below 3 0 the noise of the sensor will increase At pH 10 0 and higher the response of the sensor will diminish significantly SENSOR CARE AND MAINTENANCE Durability and H...

Страница 11: ...olution the membrane tip should first be soaked in a protease solution for several minutes to remove protein build up and then in distilled water Enzymatic detergent Enzol WPI 7363 can also be used Th...

Страница 12: ...p kit inject approximately 100 L of electrolyte directly into the new sleeve starting about half way down the sleeve and drawing the MicroFil out of the sleeve as it fills The filling process should b...

Страница 13: ...le WPI 600012 ACCESSORIES Part Number Description 5399 T Adapter Kit pkg of 3 600011 ISO HPO 2 Startup Kit 600012 Replacement Sleeves pkg of 4 100042 ISO HPO 2 Electrolyte Filling Solution 10mL 7363 4...

Страница 14: ...l interferences Calibration data set is not linear The dilution factors may be incorrect Verify the procedure used Uneven aliquots may have been used Check the pipetter calibration Sensitivity below r...

Страница 15: ...t special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or...

Страница 16: ...berg Hesson Germany Tel 49 0 6031 1602171 Fax 49 0 6031 1602180 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 18A No8 Dongfang Rd Pudong District Shanghai 200120 PR China...

Отзывы: