ABOUT THIS MANUAL

The following symbols are used in this guide:

This symbol indicates a CAUTION. Cautions warn against actions that can cause 

damage to equipment. Please read these carefully.

This symbol indicates a WARNING. Warnings alert you to actions that can cause 

personal injury or pose a physical threat. Please read these carefully.

NOTES and TIPS contain helpful information.

Fig. 1—The ISO-HPO-2 sensor works with WPI free radical analyzers to monitor hydrogen 

peroxide.

INTRODUCTION

The 

ISO-HPO-2

 is a 2.0 mm stainless steel sensor, with replaceable membrane 

sleeves (#

600012

) and an internal refillable electrolyte (#

100042

). The sensor is 

designed for use in cell cultures and similar applications.
The basic structure of the 

ISO-HPO-2

 sensor is shown in Fig. 2. It consists of an 

internal H

2

O

2

-sensing working/reference electrode combination. This electrode fits 

inside a disposable protective stainless steel sleeve (WPI #

600012

) which must contain 

fresh electrolyte (WPI #

100042

) and is separated from the external environment by a 

gas-permeable polymeric membrane covering the end of the sleeve. The other end of 

the sleeve is flanged. The locking cap is used to attach the sleeve to the probe handle.

Notes and Warnings

NOTE

: The sensor must be polarized for at least 2 hours in 0.1M PBS solution prior to use.

NOTE

: The sensor membrane is extremely delicate. Improper handling will lead to 

damage of the sensor.

NOTE

: For optimal use of each sensor and sensor membrane sleeve, the sensor must 

be stored properly. See “Sensor Care and Maintenance” on page 6. If your sensor 

was shipped wet, it includes a fluorescent sticker on the box that says, “THE SENSOR 

SLEEVE IS FILLED WITH ELECTROLYTE. IF THE SENSOR IS NOT TO BE USED WITHIN 

ONE WEEK, REMOVE THE SLEEVE, CLEAN THE SENSOR WITH DEIONIZED WATER, AND 

STORE THE SENSOR DRY.” See “Cleaning the Membrane” on page 7.

Содержание ISO-HPO-2

Страница 1: ...WORLD PRECISION INSTRUMENTS ISO HPO 2 Hydrogen Peroxide Sensor for use with WPI Analyzers Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 031919...

Страница 2: ......

Страница 3: ...king of 0 1M PBS Buffer Solution 5 SENSOR CARE AND MAINTENANCE 6 Durability and Handling 6 Storing the Sensor 7 Cleaning the Membrane 7 Replacing the Membrane Sleeve 7 ACCESSORIES 9 TROUBLESHOOTING 10...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...working reference electrode combination This electrode fits inside a disposable protective stainless steel sleeve WPI 600012 which must contain fresh electrolyte WPI 100042 and is separated from the...

Страница 6: ...if any parts are missing at 941 371 1003 or customerservice wpiinc com Returns Do not return any goods to WPI without obtaining prior approval RMA required and instructions from WPI s Returns Departme...

Страница 7: ...includes replacement membrane sleeves along with all the accessories to fill them properly with filling solution electrolyte OPERATING INSTRUCTIONS Polarizing the Sensor 1 Measure an appropriate volum...

Страница 8: ...f 450mV The oxidation of H2 O2 at the sensor surface produces a small current pA which is detected by the TBR4100 1025 The amount of current produced is linearly proportional to amount of H2 O2 in the...

Страница 9: ...er a half hour in solution it is likely that the membrane has been damaged and the sleeve needs to be changed See Replacing the Mem brane Sleeve on page 7 NOTE The calibration should be carried out at...

Страница 10: ...nsitivity However if the pH is below 3 0 the noise of the sensor will increase At pH 10 0 and higher the response of the sensor will diminish significantly SENSOR CARE AND MAINTENANCE Durability and H...

Страница 11: ...olution the membrane tip should first be soaked in a protease solution for several minutes to remove protein build up and then in distilled water Enzymatic detergent Enzol WPI 7363 can also be used Th...

Страница 12: ...p kit inject approximately 100 L of electrolyte directly into the new sleeve starting about half way down the sleeve and drawing the MicroFil out of the sleeve as it fills The filling process should b...

Страница 13: ...le WPI 600012 ACCESSORIES Part Number Description 5399 T Adapter Kit pkg of 3 600011 ISO HPO 2 Startup Kit 600012 Replacement Sleeves pkg of 4 100042 ISO HPO 2 Electrolyte Filling Solution 10mL 7363 4...

Страница 14: ...l interferences Calibration data set is not linear The dilution factors may be incorrect Verify the procedure used Uneven aliquots may have been used Check the pipetter calibration Sensitivity below r...

Страница 15: ...t special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or...

Страница 16: ...berg Hesson Germany Tel 49 0 6031 1602171 Fax 49 0 6031 1602180 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 18A No8 Dongfang Rd Pudong District Shanghai 200120 PR China...

Отзывы: