EZ Anesthesia

World Precision Instruments 

5

Water Pump And Surgical Pump Setup

1.  Connect both tubes from the surgical bed to the two water pump tubes.
2.  Before filling the pump, be sure the in/out water tubes between the pump and 

surgical bed are connected.

3.  For proper performance, the pump should be positioned at or above the surgical 

bed level, not below. Placing the pump below the bed can result in water overflow 

and water possibly leaking on to the floor, or it could reduced the motor life.

4.  Open the cap on the top of the pump. Fill the pump with room temperature 

water to just above the operating water level indicator line on the side of the 

pump and replace cap.

5.  Turn on the pump to begin circulating water to the surgical bed and choose 

appropriate temperature setting. The recommended setting for one or two beds 

is LOW. For three or more beds, the recommended setting is HIGH.

6.  The operating water level will quickly diminish. If the water level falls below the 

indicator line, turn off the pump, remove cap and add water up to the indicator 

line. Do not exceed the indicator line level. Replace the cap and operate the 

pump. Repeat steps 4 and 5 until the water stabilizes at the line level.

7.  Once the system is stabilized at the line level, run the pump continuously. The 

bed will reach the temperature set point in approximately twenty minutes.

NOTE:  Multiple surgical beds can work with one water pump.

A.  Connect pump output to first bed input.
B.  Connect first bed output to second bed input.
C.  Connect second bed output to pump input.
D.  Additional beds may be added following this basic procedure.

Isoflurane Vaporizer Setup

Pay attention to all warnings and cautions, as these draw attention to potentially 

harmful or dangerous results.

Cautions and Warnings

CAUTION

: The vaporizer must ONLY be used with Isoflurane. No other agent is 

acceptable and may be dangerous.  

CAUTION

: Isoflurane should only be used by licensed personnel. Wear 

appropriate facemask and follow safety procedures as recommended by OSHA 

and your facility safety officers. If you have any questions regarding safety, 

please contact WPI for assistance.

Содержание EZ Anesthesia

Страница 1: ...WORLD PRECISION INSTRUMENTS EZ Anesthesia Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 043019...

Страница 2: ......

Страница 3: ...ree Phases 3 Anatomy of an Anesthesia System 3 SETUP PROCEDURES 4 Water Pump And Surgical Pump Setup 5 Isoflurane Vaporizer Setup 5 Cautions and Warnings 5 Filling the Vaporizer 6 OPERATING THE SYSTEM...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...ting instructions and not following standard laboratory safety procedures when working with anesthetic agents and volatile compressed gases The system and its components must only be modified or repai...

Страница 6: ...ygen Regulator Flow Meter 1 Vaporizer Manifold Flow Meter 2 Induction Chamber Oxygen Flush Charcoal Filter Canister Charcoal Filter Canister Breathing Device Nose cone Surgical Bed Water Heating Pump...

Страница 7: ...g liquid to create an anesthetic gas at the exact concentration level required Manifold distributes anesthetic gas to the induction chamber and the breathing devices Flowmeter 2 controls the flow rate...

Страница 8: ...induction chamber input B Attach air filter canister to the white plastic filter adaptor at the end of the large diameter exhaust line coming from the chamber C Slide the air filter canister onto the...

Страница 9: ...ls below the indicator line turn off the pump remove cap and add water up to the indicator line Do not exceed the indicator line level Replace the cap and operate the pump Repeat steps 4 and 5 until t...

Страница 10: ...vaporizer upright at all times and do not carry the vaporizer by holding the dial control Filling the Vaporizer 1 The control dial must be in OFF position Press the lock button to turn the dial Alway...

Страница 11: ...g and surgical bed is at proper temperature WARNING IMPORTANT MAKE SURE THE SMALL BLACK PLUG IS INSERTED INTO BREATHING DEVICE NOSECONE THE OPERATOR COULD FACE EXPOSURE TO ISOFLURANE IF THE PLUG IS NO...

Страница 12: ...e moved to the breathing unit be sure that the desired breathing units are connected to the manifold 2 Reduce the secondary flowmeter setting to zero then press the oxygen flush button for 5 10 second...

Страница 13: ...o appropriately marked containers and disposed of properly Less frequent intervals may be used when the anesthetic agent does not contain additives or a stabilizing agent CAUTION Do not drain the vapo...

Страница 14: ...e top of the water pump and tip the pump over a sink until all the water is emptied from the pump 3 Replace the cap on the pump Disinfecting the Water Pump and Surgical Bed The water pump and surgical...

Страница 15: ...t special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or...

Страница 16: ...berg Hesson Germany Tel 49 0 6031 1602171 Fax 49 0 6031 1602180 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 18A No8 Dongfang Rd Pudong District Shanghai 200120 PR China...

Отзывы: