Shutting Down And Flushing The System
1. Turn the vaporizer dial to the OFF position, making sure it is locked. Allow the
oxygen to flow for 2 minutes to flush the gas from the host cages and breathing
chambers.
2. Shut off the oxygen tank.
3. Allow the oxygen to flow until the pressure has been bled from the tank.
4. Flush the line:
•
EZ-7000/EZ-150C; EZ-FF9000/EZ-190F
– Push the green oxygen flush button
to be sure no pressure remains in the flush line.
•
EZ-B800/EZ-108SA
– Turn the chamber lever to CHAMBER FLUSH to be sure
no pressure remains in the flush line.
5. Turn the water pump switch to the OFF position.
6. To drain the fluid system, see Cleaning and Maintenance procedures (page 9).
CLEANING AND MAINTENANCE PROCEDURES
Draining the Vaporizer
Every two weeks, the vaporizer should be drained into appropriately marked
containers and disposed of properly. Less frequent intervals may be used when the
anesthetic agent does not contain additives or a stabilizing agent.
CAUTION
: Do not drain the vaporizer into a container that is not properly
marked.
1. Set the control dial in the OFF position.
2. Remove the filler cap from the fill cylinder to reveal the bottom drain plug.
3. Place a suitable container under the drain outlet under the fill cylinder.
4. Slowly unscrew the drain plug until the liquid begins to flow into the container.
DO NOT completely unscrew the drain plug.
5. When all of the liquid agent has been drained, tighten the drain plug and replace
the filler cap.
Servicing the Vaporizer
The vaporizer must be fully serviced and calibrated at least every three years. Servicing
must be done by an authorized service technician. Contact WPI at 941.371.1003 or
[email protected] for instructions
CAUTION
: DO NOT MODIFY, TAMPER WITH OR DISASSEMBLE THE VAPORIZER.
THIS UNIT MUST ONLY BE SERVICED BY AUTHORIZED PERSONNEL.