background image

World Precision Instruments

 WARNING: PREGNANT WOMEN SHOULD NOT BE EXPOSED TO 
ISOFLURANE.

CAUTION

: During use, frequently check that the Isoflurane level is between the 

minimum and maximum marks on the sight glass level indicator. Refill the 

vaporizer before the liquid level reaches the minimum mark.

CAUTION

: The vaporizer must NEVER be modified, dismantled, calibrated or 

serviced by unauthorized personnel. The vaporizer MUST be serviced at an 

approved Service Center.  The unit should be serviced for cleaning and 

calibration every one to two years.

CAUTION

: The vaporizer MUST be connected so that the flow of gas to the 

animal is as indicated by the arrows on the device. The incorrect flow direction 

will result in wrong dosage.

CAUTION

: Before use, ALL connections must be checked for leaks and 

functional tests MUST be performed, as is the normal procedure for an 

anesthetic device.

CAUTION

: Keep the vaporizer upright at all times and do not carry the 

vaporizer by holding the dial control.

Filling the Vaporizer

1.  The control dial must be in OFF position. Press the lock button to turn the dial. 

Always keep the dial in the OFF position when the unit is not in use.

2.  Verify that the drain plug is closed.

3.  Remove the filler cap by turning it counterclockwise.

4.  Verify that the liquid agent is Isoflurane.

5.  Slowly pour the agent into the opening.

6.  Observe the proper level through the sight glass on front of the vaporizer.

7.  Replace the filler cap and tighten it securely.

OPERATING THE SYSTEM

 

CAUTION

: Before using the system, verify that all elements are set up and 

functioning properly, following safety checks as recommended by OSHA and 

your safety officer. Check the system for gas leaks. Be sure that the water 

tubes and fittings are not leaking.

WARNING

:  HIGH ISOFLURANE TO OXYGEN RATIO CAN RESULT IN 

HARM OR DEATH TO ANIMALS.

IMPORTANT: The following Isoflurane flow and mix rates are a suggested starting point 

for developing your protocols. All settings should be refined and adjusted based on 

Содержание EZ Anesthesia

Страница 1: ...WORLD PRECISION INSTRUMENTS EZ Anesthesia Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 043019...

Страница 2: ......

Страница 3: ...ree Phases 3 Anatomy of an Anesthesia System 3 SETUP PROCEDURES 4 Water Pump And Surgical Pump Setup 5 Isoflurane Vaporizer Setup 5 Cautions and Warnings 5 Filling the Vaporizer 6 OPERATING THE SYSTEM...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...ting instructions and not following standard laboratory safety procedures when working with anesthetic agents and volatile compressed gases The system and its components must only be modified or repai...

Страница 6: ...ygen Regulator Flow Meter 1 Vaporizer Manifold Flow Meter 2 Induction Chamber Oxygen Flush Charcoal Filter Canister Charcoal Filter Canister Breathing Device Nose cone Surgical Bed Water Heating Pump...

Страница 7: ...g liquid to create an anesthetic gas at the exact concentration level required Manifold distributes anesthetic gas to the induction chamber and the breathing devices Flowmeter 2 controls the flow rate...

Страница 8: ...induction chamber input B Attach air filter canister to the white plastic filter adaptor at the end of the large diameter exhaust line coming from the chamber C Slide the air filter canister onto the...

Страница 9: ...ls below the indicator line turn off the pump remove cap and add water up to the indicator line Do not exceed the indicator line level Replace the cap and operate the pump Repeat steps 4 and 5 until t...

Страница 10: ...vaporizer upright at all times and do not carry the vaporizer by holding the dial control Filling the Vaporizer 1 The control dial must be in OFF position Press the lock button to turn the dial Alway...

Страница 11: ...g and surgical bed is at proper temperature WARNING IMPORTANT MAKE SURE THE SMALL BLACK PLUG IS INSERTED INTO BREATHING DEVICE NOSECONE THE OPERATOR COULD FACE EXPOSURE TO ISOFLURANE IF THE PLUG IS NO...

Страница 12: ...e moved to the breathing unit be sure that the desired breathing units are connected to the manifold 2 Reduce the secondary flowmeter setting to zero then press the oxygen flush button for 5 10 second...

Страница 13: ...o appropriately marked containers and disposed of properly Less frequent intervals may be used when the anesthetic agent does not contain additives or a stabilizing agent CAUTION Do not drain the vapo...

Страница 14: ...e top of the water pump and tip the pump over a sink until all the water is emptied from the pump 3 Replace the cap on the pump Disinfecting the Water Pump and Surgical Bed The water pump and surgical...

Страница 15: ...t special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or...

Страница 16: ...berg Hesson Germany Tel 49 0 6031 1602171 Fax 49 0 6031 1602180 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 18A No8 Dongfang Rd Pudong District Shanghai 200120 PR China...

Отзывы: