background image

EZ Anesthesia

World Precision Instruments 

7

your own lab testing. Factors specific to your laboratory context may impact protocols. 

Protocols should be developed based on your requirements and careful observation 

of animals during anesthesia. Check the reflexes of the animal before transferring 

them to the surgical table.

Prior to Operation

WARNING: THE SYSTEM MUST BE PROPERLY SET UP BEFORE TURNING 
ON ANY GAS.

NEVER disconnect any hoses until the oxygen has been shut off and the vaporizer 

is shut off. Failure to precisely follow these instructions could result in injury. Back 

pressure can cause the vaporizer hoses to pop, and it can cause Isoflurane to backup 

and leak into the manifold and possibly the flowmeter, therefore affecting system 

performance and accuracy.
1.  Verify that all tubes are properly connected.
2.  Check that water pump is operating and surgical bed is at proper temperature.

WARNING: IMPORTANT! MAKE SURE THE SMALL BLACK PLUG IS 

INSERTED INTO BREATHING DEVICE NOSECONE. THE OPERATOR 

COULD FACE EXPOSURE TO ISOFLURANE IF THE PLUG IS NOT INSERTED 

PROPERLY.

3.  Check that the inlet flowmeter on the vaporizer stand is fully closed by turning it 

clockwise.

4.  Connect the desired components to be used with the system (for example, 

breathing devices and chamber). At least one component must be connected 

before turning on any gas.

5.  If the induction chamber is to be used for initial anesthesia, verify that the 

secondary flow meter is turned fully counterclockwise to open.

7.  Fully open the oxygen tank valve.
8.  Turn the inlet flowmeter knob counterclockwise to open and adjust it to:

• 

Mice: 1 liter/min.

• 

Rats: 1.5 liter/min.

• 

Guinea Pigs: 3 liter/min.

9.  If you are using the induction chamber, check for oxygen flow through the 

secondary flowmeter. The secondary flow meter should read between 1-3 liters/

Min. depending on the species.

Induction

1.  Place animals into the induction chamber and latch the lid.

Содержание EZ Anesthesia

Страница 1: ...WORLD PRECISION INSTRUMENTS EZ Anesthesia Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 043019...

Страница 2: ......

Страница 3: ...ree Phases 3 Anatomy of an Anesthesia System 3 SETUP PROCEDURES 4 Water Pump And Surgical Pump Setup 5 Isoflurane Vaporizer Setup 5 Cautions and Warnings 5 Filling the Vaporizer 6 OPERATING THE SYSTEM...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...ting instructions and not following standard laboratory safety procedures when working with anesthetic agents and volatile compressed gases The system and its components must only be modified or repai...

Страница 6: ...ygen Regulator Flow Meter 1 Vaporizer Manifold Flow Meter 2 Induction Chamber Oxygen Flush Charcoal Filter Canister Charcoal Filter Canister Breathing Device Nose cone Surgical Bed Water Heating Pump...

Страница 7: ...g liquid to create an anesthetic gas at the exact concentration level required Manifold distributes anesthetic gas to the induction chamber and the breathing devices Flowmeter 2 controls the flow rate...

Страница 8: ...induction chamber input B Attach air filter canister to the white plastic filter adaptor at the end of the large diameter exhaust line coming from the chamber C Slide the air filter canister onto the...

Страница 9: ...ls below the indicator line turn off the pump remove cap and add water up to the indicator line Do not exceed the indicator line level Replace the cap and operate the pump Repeat steps 4 and 5 until t...

Страница 10: ...vaporizer upright at all times and do not carry the vaporizer by holding the dial control Filling the Vaporizer 1 The control dial must be in OFF position Press the lock button to turn the dial Alway...

Страница 11: ...g and surgical bed is at proper temperature WARNING IMPORTANT MAKE SURE THE SMALL BLACK PLUG IS INSERTED INTO BREATHING DEVICE NOSECONE THE OPERATOR COULD FACE EXPOSURE TO ISOFLURANE IF THE PLUG IS NO...

Страница 12: ...e moved to the breathing unit be sure that the desired breathing units are connected to the manifold 2 Reduce the secondary flowmeter setting to zero then press the oxygen flush button for 5 10 second...

Страница 13: ...o appropriately marked containers and disposed of properly Less frequent intervals may be used when the anesthetic agent does not contain additives or a stabilizing agent CAUTION Do not drain the vapo...

Страница 14: ...e top of the water pump and tip the pump over a sink until all the water is emptied from the pump 3 Replace the cap on the pump Disinfecting the Water Pump and Surgical Bed The water pump and surgical...

Страница 15: ...t special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or...

Страница 16: ...berg Hesson Germany Tel 49 0 6031 1602171 Fax 49 0 6031 1602180 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 18A No8 Dongfang Rd Pudong District Shanghai 200120 PR China...

Отзывы: